Procedure Library
P17.5

Density and Specific Gravity — Hydrometer, Digital Densitometer, and Pycnometer Methods

Procedure 5 of 40 · Chapter 17 · §17.1.6

1 table 0 notes

Table 1

FeatureMethod A: HydrometerMethod B: Digital DensitometerMethod C: Pycnometer
StandardASTM D1298, ISO 3675ASTM D4052, ISO 12185ASTM D1217, ISO 3507
Sample volume500 mL1–2 mL10–50 mL
Precision (repeatability)±0.0005 g/cm³±0.0001 g/cm³±0.00002 g/cm³
Temperature control±0.25°C water bath±0.02°C internal Peltier±0.1°C water bath
Measurement time15–30 min1–3 min30–60 min
Viscosity limit15 000 mm²/s15 000 mm²/s5 000 mm²/s
Operator skillModerateLow (automated)High
Capital costLowHighLow